Description

For USB operation
Key facts
  • 10 000 triggered measurements/s
  • More than 50 000 readings/s
  • Unprecedented measurement speed and accuracy even at low levels
  • Ideal for universal applications
  • Detachable cables for flexible operation
  • 93 dB dynamic range thanks to innovative three-path concept
  • Built-in trigger I/O port

Brief description The R&S®NRP18S-20 high-power sensor consist of an R&S®NRP18S and a 20 dB attenuator and is able to perform power measurements up to 15 W.
Specifications
High-power three-path diode power sensor
Frequency range Power measurement range Impedance matching (SWR) Rise time video bandwidth Uncertainty for power measure- ments at +20 °C to +25 °C absolute (in dB) Uncertainty for power measure- ments at +20 °C to +25 °C relative (in dB)
R&S®NRP18S-10, N (m)
10 MHz to 18 GHz 1 nW to 2 W (–60 dBm to +33 dBm) 10 MHz to 2.4 GHz: < 1.14 <5 µs> 100 kHz 0.083 to 0.198 0.022 to 0.087
    >2.4 GHz to 8.0 GHz: < 1.20      
    > 8.0 GHz to 12.4 GHz: < 1.25      
    > 12.4 GHz to 18.0 GHz: < 1.30      
R&S®NRP18S-20, N (m)
10 MHz to 18 GHz 10 nW to 15 W (–50 dBm to +42 dBm) 10 MHz to 2.4 GHz: < 1.14 <5 µs> 100 kHz 0.083 to 0.198 0.022 to 0.087
    >2.4 GHz to 8.0 GHz: < 1.25      
    > 8.0 GHz to 12.4 GHz: < 1.30      
    > 12.4 GHz to 18.0 GHz: < 1.41      
R&S®NRP18S-25, N (m)
10 MHz to 18 GHz 30 nW to 30 W (–45 dBm to +45 dBm) 10 MHz to 2.4 GHz: < 1.14 <5 µs> 100 kHz 0.083 to 0.198 0.022 to 0.087
    >2.4 GHz to 8.0 GHz: < 1.25      
    > 8.0 GHz to 12.4 GHz: < 1.30      
    > 12.4 GHz to 18.0 GHz: < 1.41      
''